6. TECHNICAL DATA

6.1 Photometric data

Central illumination intensity
at a distance of 1 meter
Light spot diameter d
10
Light spot diameter d
50
Residual illumination intensity with
one shutter
Residual illumination intensity with
two shutters
Residual illumination intensity at the
bottom of a normed tube
Residual illumination intensity at
the bottom of a normed tube and
one shutter
Residual illumination intensity at the
bottom of a normed tube and two
shutters
Illumination depth 20 %
Illumination depth 60 %
Colour rendering index R
a
Colour rendering index R
9
Max. irradiation intensity in the field
at a distance of 1.0 m
Max. irradiation intensity in the field
at a distance of 0.8 m
Focusable light spot size
Colour temperature (Kelvin)
Heat build-up near the head
Electronic brightness control on the
light body (Standard)
Number of LEDs
Working lifespan of the LEDs
Work area
Diameter of the light body
Height adjustment limit
Note: The technical data are subject to certain deviations.
Due to production–related reasons, the actual values can
vary slightly from the above-named values. The values for R
can deviate by approx. ± 5%. The colour temperature values
can deviate by approx. ± 200K.

6.2 Electrical Data

18W, Focus
18W
45,000 Lux
35,000 Lux
122 mm
132 mm
62 mm
66 mm

6.3 Electrical installation information

The KaWe examination light is exposed to a current peak when it is
0 %
0 %
switched on and is therefore equipped with a power adapter. The
adapter has a wide range input, input voltage 100 - 240V AC, 50 -
60 %
61 %
60Hz, Output voltage 24V DC.
If a changeover relay is required for an existing on-site emergency
100 %
100 %
power supply for the KaWe adapter, this relay must be ordered sep-
arately from KaWe.
0 %
0 %
Warning! This light is a Protection Class I device. To avoid
the risk of electric shock, this unit must be connected to a
60 %
61 %
grounded electric outlet.

6.4 Weights

1750 mm
1750 mm
890 mm
840 mm
95
95
94
94
162 W/m
2
126 W/m
2

6.5 Environmental conditions

201 W/m
148 W/m
2
2
14-20 cm
17 cm
4300 K
4300 K
0.5° C
0.5° C
50 - 100 %
50 - 100 %
* for higher temperatures please contact us
12
12
50,000 h
50,000 h
70-140 cm
70-140 cm
29 cm
29 cm
121 cm
121 cm
Instructions on the packaging
a
Power input
Operating voltage
Amperage
KaWe examination lamp
18W
18W, Focus
Temperature
Relative humidity
Air pressure
Temperature
Relative humidity
Air pressure
Temperature range
Humidity during
during transport
transport and
and storage
storage
+50° C
RH
20% - 90%
–10° C
18W, Focus / 18W
18 W
24 V DC
0.75 A
Weight
1.9 kg
1.9 kg
Operation
Min.
Max.
+10° C
+30° C*
30 %
75 %
700 hPa
1060 hPa
Transport/storage
Min.
Max.
-10° C
+50° C
20 %
90 %
700 hPa
1060 hPa
Air pressure during
transport and
storage
P
700hPa - 1060hPa