Fault cause
No gas flow
OPSO cannot be unlocked
Restarting the safety overpressure shut-off valve OPSO with visual indicator
If the safety overpressure shut-off valve OPSO has responded, which is shown in RED on
the visual indicator - take the following steps.
1. Close the gas withdrawal valve.
2. Vent the pressure regulator loosen the screw connection on the outlet side
3. Ventilate the inlet pressure Loosen the connection on the inlet side
4. Tighten all connections again.
5. After elimination of defects, open the gas cylinder valve or the shut-off fittings.
6. Unscrew the protective cap
7. Turn the protective cap
until the spindle
8. Screw the protective cap
9. OPSO
is ready to operate and the display is GREEN.
10. Check for leaks.
10 / 12
Low pressure regulator Type NDR 0515 and Type NDR 0516
Action
Gas feed is closed:
open the gas cylinder valve or the shut-off fittings
Safety overpressure shut-off valve (OPSO / UPSO) is closed:
carry out the steps described in "Restarting the safety
overpressure shut-off valve (OPSO / UPSO)"
Filter sieve in the input connection is dirty:
send the pressure regulator to the manufacturer to be
checked
The thermal cut-out device has responded.
Replace the pressure regulator.
Regulator membrane is damaged:
Pressure regulator is defective, replace.
by hand.
around and pull out the spindle
clicks into place and remains open.
on again by hand.
.
with the release device
.