Communication

5.1

The Serial Interface Of The Transducer

The transducer VSP/VCP is equipped with a serial RS485 interface. To set the
device address for communication via RS485 please remove the rubber cap over
the address switch (1) and then set the address switch to a value between 1 and
16 using a small screw driver or a similar tool (2). Afterwards insert the rubber
cap again.
Communication telegrams are transmitted as ASCII text according to the
Thyracont protocol version V2. Detailed information is provided in the separate
description "Thyracont Communication Protocol".
Download link:
www.thyracont-vacuum.com/download-center/
The analog output signal 0-10V is simultaneously available with
RS485 !
Interface-Parameter:
9,6 / 14,4 / 19,2 / 38,4 / 57,6 / 115,2 kBd, 8 data bits, 1 Stopbit, no parity
When powered on the transducer starts with 9,6 kBd. If a tele-
gram with different baud rate is received, the transducer will au-
tomatically adapt to it. For this automatic baud rate adaption a
maximum of two telegrams of type "Type Query" or "Measure-
ment Query " is required.