ABOUT OPTIFLEX 3 CPM UNIT
U.S. Federal law restricts this device to sale by, or on the order of, a
physician or licensed practitioner.
Make certain that the unit is electrically grounded by connecting
only to a grounded electrical service receptacle conforming to the
applicable national and local electrical codes.
Keep hair, loose clothing, loose bedding, fingers and toes away
from the hinge components of the unit.
Do not use the OptiFlex 3 outdoors or on wet surfaces. Use only on
firm, flat, stable level surfaces to ensure stability of the unit while in
operation.
Materials of the unit may become flammable or combustible if
exposed to a source of ignition.
Heat generated within the pendant may cause ignition of the
pendant if wrapped in bedding or other materials.
Do not use OptiFlex 3 while smoking or around open flame.
OptiFlex 3 has been designed for protection against the exposure
of urinary incontinence. Precautionary measures should still be
taken when any type of liquid comes in contact with an electrical
apparatus.
Always turn off and unplug unit from electrical source before
servicing or cleaning. Failure to do so could result in electrical
shock or personal injury.
Handle the unit only when unit is dry and hands are dry to prevent
electrical shock.
Do not use the OptiFlex 3 as a toy.
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
Exercise caution when using accessories and auxiliary devices such
as muscle stimulators, cold packs and other modalities. Route lead
wires, hoses, tubes, etc. away from the working mechanism of the
OptiFlex 3 to help prevent damage to the Optiflex 3 and any other
modality used with it.
Unconscious patients or patients under heavy influence of
medication must be constantly attended and monitored while the
OptiFlex 3 is in use.
The OptiFlex 3 unit must be completely visible at all times during
use. Never cover the unit with bedding or any other means of
concealment while in operation.
If the OptiFlex 3 is used in conjunction with the optional OptiFlex
"T" trolley, make certain the OptiFlex 3 unit is resting on the
mattress of the bed and the OptiFlex "T" is suspended with no
weight on the casters to prevent possible movement of the unit
and or possible injury to patient.
4