NOMENCLATURE
PENDANT FUNCTIONS AND USE
Start/Stop (Emergency Stop): Starts or stops the
treatment.
How to Use: Press the Start/Stop (Emergency Stop)
button to begin or end the treatment.
NOTE: When unit is stopped and the Start/Stop
(Emergency Stop) button is pressed, the unit will
immediately begin moving in the opposite direction of the
last movement.
Comfort Zone ™ : Temporarily decreases the flexion
ROM when discomfort is experienced during treatment.
How to Use: Press and the carriage will immediately
reverse direction to move toward the extension angle.
For the next five cycles, the carriage will operate in a
range 5° less than the angle where the button was
pressed. The Comfort Zone Flexion Angle is 5° less than
the angle where the button was pressed. Once five
cycles are complete, the Comfort Zone Flexion Angle
will be increased by 1° each consecutive cycle until it
reaches the currently Programmed Flexion Angle. See
Chart 1.
NOTE: If Comfort Zone is pressed within 30° of the
currently Programmed Extension Angle, treatment will
stop and the message "Comfort Zone Too Small" will be
displayed.
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
CHART 1 COMFORT ZONE
C = Comfort Zone activated temporary flexion limit established.
5° = 5° less than the angle where Comfort Zone was activated.
Upon completion of five cycles, ROM will increase by 1° each cycle until it reaches P.
P = Programmed Flexion angle.
S = Start/Finish.
If pain is severe, do not use the Comfort Zone function. Stop
treatment immediately. Failure to stop treatment could result in
tissue damage and could compromise the surgical repair.