NOMENCLATURE
PENDANT FUNCTIONS AND USE
Mode: Used together with Fast Back ™ , Oscillation
Zone ™ and Progressive ROM button to enable and
disable the appropriate mode. All three modes or any
combination may be used at once.
How to Use: Press Mode button along with desired
function (Fast Back, Oscillation Zone and/or
Progressive ROM).
Fast Back ™ : When the carriage reaches 15° less
than the Programmed Flexion Angle, it will begin to
slow down. The deceleration rate will be half of the
Programmed Speed when the carriage comes within
10° of the flexion angle. Upon reaching the flexion
angle, the carriage will reverse direction and resume the
Programmed Speed as it moves towards the extension
angle. See Chart 2.
How to Use: While pressing and holding the Mode
button, press the Fast Back button to enable the Fast
Back mode.
NOTE: If both Fast Back and Oscillation Zone modes
are enabled, the carriage will move at the slower of 75°/
min. and half of the Programmed Speed while in the
Oscillation Zone ROM.
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
CHART 2 FAST BACK
Non-Working ROM
Fast Speed
Fast Speed
F = Programmed Flexion Angle
E = Programmed Extension Angle
9
Working ROM
(10°)
Slow Speed
Fast Speed
Working ROM
Speed = 60°/
min.