Roller 6/10 digital
®
The AC adapter is suitable for 100 V upto 240 V. Insert an appro-
priate adapter corresponding to the country, before plugging the
AC adapter into the mains socket.
Switching the device on
Press the "Power" button (L) to switch the device on.
The following data appear on the display in the sequence listed:
the test image
the software version
the operating mode
the values and settings last used
Speed setting
Use the "Speed setting" button (K) to set the speed to a value
between 5 and 80 rpm in units of 1 rpm.
Timer/Counter
Select "Counter" or "Timer" (I).
Timer
When "Timer" is selected, the set value counts down. Once the
timed shaking cycle has been completed the device switches off
automatically.
 An acoustic signal beeps three times and the "Timer" LED (H)
flashes red.
Counter
The counter measures the time from the start of shaking to the
end.
 The "Counter" LED (G) flashes red.
If the "Start/Stop" button (F) is pressed the shaking process stops
and the elapsed time is displayed. If the counter is restarted the
measurement restarts afresh.
To switch from counter to timer or vice versa, press the "Timer"
button (I).
Setting
Set the timer from 1 sec to 99 h 59 min using the "Plus/Minus"
timer setting buttons (J).
Press the "Start/Stop" button (F) to start the shaker.
The last timer value setting is saved and is loaded for the next
shaking process.
Resetting to 0
To reset the timer, press the "Timer Minus" and "Timer Plus" but-
tons (J) simultaneously.
Starting the device
Press the "Start/Stop" button (F) to start the shaking process.
The device starts with the desired settings.
Stopping the device
Press the "Start/Stop" button (F) to pause operation.
Switching off the device
Press the "Power" button (L) to switch the device off.