Error symptom
The sensor obviously
supplies incorrect
measured values
The sensor cannot be
calibrated
The sensor still cannot
be calibrated after
cleaning
Incorrect temperature
display
Mechanical damage to
the measuring tip
The battery LED flashes – The batteries are almost empty
6
What to do if...
Cause
– No connection between
measuring instrument and sensor
– Cable defective
– Sensor switched off
– pH sensitive area and/or junction
contaminated
– Zero adjustment erroneous
– Sensor not sufficiently watered
– Measuring tip depleted
– Temperature sensor defective
– Cable defective
Remedy
– Check connection between
measuring instrument and sensor
– Replace the sensor
– Switch on the sensor
– Clean the measuring tip (see
section 5.9)
– Carry out zero adjustment
– Watering the sensor
– Exchange the measuring tip (see
section 5.10)
– Exchange the measuring tip (see
section 5.10)
– Replace the sensor
– Exchange the measuring tip (see
section 5.10)
– Insert new batteries (see
section 3.2)