Zero adjustment and calibration

Before calibrating in pH buffer solutions, the measuring tip has to be watered
and the zero point of the sensor has to be adjusted (zero adjustment).
l Place the sensor with the measuring tip in a vessel with 3 mol/l KCl
solution for at least 10 minutes. The pH sensitive area and the junction
must be completely submersed in the solution while doing so.
l Subsequently, thoroughly rinse the measuring tip with deionized water
and shake off the water.
Note
If you do not have any 3 mol/l KCl solution you can also water the measuring
tip in pH 7 buffer solution. For precise measurements, however, we
recommend to use 3 mol/l KCl solution (solution see section 5 M
,
CLEANING