Measuring principle
Temperature compensation
Meas. conditions
Meas. range
Temperature range
Max. admissible overpressure
Immersion depth
Operating position
Incident flow
Characteristics when delivered
Zero signal
Response time at 20°C
Self consumption at 20°C
Temperature adaptation
Drift
Operation time
Material
Dimensions
62
Technical data
Membrane covered galvanic sensor
IMT
0 - 50 mg/l O 2
0°C - 50°C
6 bar
min.
6 cm
max. 20 m depth
any
> 3 cm/s at
10 % meas. accuracy
10 cm/s at
5 % meas. accuracy
18 cm/s at
1 % meas. accuracy
< 0.1 % from saturation value
t 90 (90 % of indication of end value after) < 10 s
t 95 (95 % of indication of end value after) < 16 s
t 99 (99 % of indication of end value after) < 60 s
0.008 µg h -1 (mg/l) -1
IMT-compensation
approx. 3 %/month in operation condition
at least 6 months per electrolyte filling
Membrane head
Membrane
Sensor head
Thermistor housing
Shaft
145 mm
15.25 mm
13 µm
1.5 m bis 20 m
BA22204/06.00/Pla/Oxi323-325-SET-2
POM
FEP
POM
VA-steel 1.4571
POM
shaft length
shaft diameter
membrane thickness
cable length