Herunterladen Inhalt Inhalt Diese Seite drucken

wtw pH 526 Bedienungsanleitungen Seite 44

Mikroprozessor ph-mv-meter
Inhaltsverzeichnis

Werbung

Data storage
Display
Keyboard
Inputs
combined electrode or
measuring electrode
temperature sensor
socket for power supply
Outputs
analog output
44
Technical data
Ring memory
Number of values: 50 pairs
Output:
Data protection:
WTW-specific LCD display
height: 35 mm
breadth: 60 mm
Field with 10 keys
plug acc. to DIN 19262
WTW plug acc. to IP 66
input resistance: > 1 ⋅ ⋅ 10
< 1 ⋅ ⋅ 10
input current:
4 mm jack
TFK 150 as separated probe or NTC integrated
in pH electrode
Pt 1000 as separated probe or integrated in pH-
electrode
automatic detection of connected probe type
manual temperature input with not connected
temperature (input range: - 20°C ... + 130°C)
2-pin special socket for line adaptor
voltage: 12 V
current: 70 mA
2 x 4 mm jack for recorder,
not galvanically isolated
output signal corresponds to
electrode voltage,
i.e. 50 - 62 mV/pH-unit
Accuracy:
± 0.1 % of measured value ± 4 mV at
ambient temperature -10°C ... +55°C
Internal resistance:
< 5 Ohm (current limit max. 0.2 mA
Display,
RS 232 interface
any long
12
Ohm
-12
A
offset current)

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ph 538

Inhaltsverzeichnis