10
Menu Access Switch
The menu access switch, located on the back panel of the terminal, is covered
by a protective cap.
After configuring the operating menu and the A/D converter for the connected
load cells or analog weighing platforms, close the menu access switch to protect
menu settings and then turn the terminal off (and back on again) to reset the
software.
§ Use a screwdriver to remove the cap.
(Note: if the switch is covered by a threaded cap, turn the cap in the
required direction to remove it.)
§ Move the menu access switch to the left to lock the menu.
§ Replace the cap.
§ Turn the terminal off and back on again.
Warmup Time
To deliver exact results, the terminal must warm up for at least 30 minutes
after initial connection to AC power or after a relatively long power outage.
Only after this time will the equipment have reached the required operating
temperature. Following initial set-up, a warmup time of 10 to 20 hours
is recommended.
Note:
Configuration data is stored in battery-backed memory. Data is saved for
approximately 3 months with the FCT01-X disconnected from power. If you plan
to leave the FCT01-X disconnected from power for an extended period, we
recommend printing out the configuration data beforehand.