5.1 Removal of chuck
Open the chuck 23 as
described in section
4.1.2. Insert a tool or
the included test drill
19. Now you place the
included maintenance tool 24 on the chuck and un -
screw it anti-clock wise. Clean the chuck thoroughly
with compressed air or in an ultrasonic bath. Sub -
sequently reinsert the chuck and tighten it.
5.2 Separation of crank and motor
The procedure is described
for right-handers. Lefthan -
ders please proceed diame-
trically. Take the handpiece
in your right hand. The
motor 20 is thus in the palm of your right hand. Take
the crank 18 in your left hand and turn the crank
clockwise until you feel a marked resistance. Now
turn the crank further until motor and handpiece
separate from each other. For the assembly of the
two parts you proceed in reversed order and screw
crank and motor firmly together. Turn the crank a
quarter revolution backwards to close the chuck.
5.3 Change of motor cable
The plug of the motor cable only fits in one position
on the motor. Do not try to force the plug on the
motor. This could damage the connector pins and
makes the motor unserviceable. The procedure is
described for right-handers. Lefthanders please pro-
ceed diametrically.
Take the handpiece back-
wards in your left hand. The
motor thus is in the palm of
your left hand. Unscrew
with your right hand the
protective sleeve anti-clockwise. Carefully pull the
motor cable. The plug connection comes loose. Put
the new motor cable carefully on the motor and
screw the protective sleeve clockwise tightly on.
6
5.4 Adjustment of magnetic sensor
of knee-control plate
The speed is leadlessly controlled through a magne-
tic sensor on the knee-control plate. It may happen
that the sensor must be adjusted from time to time.
Switch the control unit off. Press the hand/foot
switch 12 and the direction selector 13 simultaneously
and keep both switches pressed. Now switch the unit
on. After approx. 2 seconds you will hear an acoustic
signal and the display shows one of the three options
Hi, Lo or rc. Turn the speed control 11 until the display
shows Hi. Now push the knee-control plate comple-
tely in and keep it there. Acknowledge with the
direction selector 13. You will hear an acoustic
signal. The maximum knee-control plate position has
been stored. Turn the speed control 11 now until the
display shows Lo. The knee-control plate must not
be pressed. Acknowledge with the direction selector
13. You will hear an acoustic signal. The minimum
knee-control plate position has been stored. You
can finish the adjustment now.
Turn the speed control 11 now until the display shows
rc and acknowledge with the direction selector 13.
Switch the control unit off and subsequently on
again. The unit is now ready for operation.
5.5 Testing of individual components
Here the function of the different components can be
tested. To load the respective menu, switch the unit
off. Press the direction selector 13 and the motor
switch 14 now and keep them pressed down. Switch
the unit on. After approx. 2 seconds you will hear an
acoustic signal. By turning the speed control 11 you
can choose from the different options. You can only
choose the option Sc after the lock button on the
speed control has been pressed. You start the test
routine by pushing the direction selector 13. You end
the test routine by again pushing the direction
selector 13.
Test setting
Function

Test procedure

Sc
Switch
Test the function of the switch and its control light
by pressing the switches 12, 13 and 14.
dc
Display
All segments of the display are switched on one after the other.
Hc
Motor sensor
Turn the motor shaft to test the function of the motor sensors.
The display shows two bars in different heights.
Fc
Foot switch
Press the foot switch. The display shows 0-9 and A-F.
nc
Knee control plate
Press the knee-control plate. The display shows 0-9 and A-F.

5.6 Replacement of unit fuse

Remove the fuse holder using a flat blade driver.
Replace the damaged fuse with the spare fuse,
which is inside the fuse holder. Put the fuse holder
back. Order a new 250 V/3,15 A fuse.

6. Storage

Store thee unit in a dry and dust-free place.

7. Technical data

Control unit:
Voltage:
220 - 240 V; 50/60 Hz
Power consumption:
40 VA
Exterior dimensions: W 94 x D 268 x H 228 mm
Weight:
2,9 kg
Handpiece:
Dimensions:
L 159 mm, ø 27 mm
Speed:
1.000 - 50.000 rpm
Cable length:
1,4 m
Weight:
230 g (without cable)
Foot switch:
Cable length:
1,8 m
Weight:
384 g
Tabletop stand:
Dimensions:
W 122 x D 268 x H 30 mm
Weight:
152 g
Handpiece tray:
Dimenstions:
W 60 x D 110 x H 45 mm
Weight:
68 g
ENU

8. Delivery includes

Control unit, Tabletop stand, Motor hand piece,
Motor handpiece tray, Foot switch, Attachment bra-
cket for control unit, Main tenance tools, Test drill.

9. Guarantee

Our technical recommendations of application are
based on our own experiences and tests and should
only be regarded as guidelines. It rests with the skills
and experience of the user to verify that the products
supplied by us are suitable for the intended proce-
dures. Our pro ducts are undergoing a continuous
further development. We reserve the right of chan -
ges in construction and composition. It is under -
stood that we guarantee the impeccable quality of
our products.
7