6. Commissioning, operation
Phase
Description
13
Delay time at the maximum set pressure
15
Time to control up to the next step
16
Delay time at the zero point
18
The delay times can be defined in the configuration.
Between phases 10 and 16, the calibrator data can be evaluated and recorded if a stable
value is displayed. The AUTO mode can be exited by pressing the MENU button. VENT
will be shown In the top left of the display.
6.4

Leakage (LEAK)

In CTRL mode, the connected test item can be checked for leakage via LEAK button.
Setting
In order to be able to check a test item for leakage, the instrument must first be placed in
CTRL mode.
1. Set the desired pressure using the
buttons.
2. As soon as the pressure is reached and the value is stable, press the LEAK button.
The internal pump will be switched off and the applied pressure will be held.
Operation
WIKA operating instructions, model CPC2000
/
button or the
/
Set pressure demand
Pressure unit
Measured value
Actual deviation from the set
value
Percentage deviation from the set
value
Elapsed time since start of the
Leak function
Operating mode
button (0 %/100 %)
23