Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo HA405-DPA-SC Betriebsanleitung Seite 9

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 6
HA405-DPA-SC pH Electrodes
5
Troubleshooting
No response (symptom)
– Poor meter connection
– Cracked glass
Slow response (symptom)
– Dehydrated pH membrane
– Clogged reference junction
– Protein contamination
– Depleted gel layer
Noise (symptom)
– Poor meter connection
– Clogged reference junction
Low slope (symptom)
– Dehydrated pH membrane
– Contaminated membrane
For further assistance, call our Customer Service Center at
+1-800-352-8763.
HA405-DPA-SC pH Electrodes
Printed in Switzerland
C heck connectors at the meter and at
the electrode
Replace electrode
Rehydrate (see Maintenance)
Soak in warm KCl (see Maintenance)
U se pH Electrode Cleaner (see Main-
tenance)
U se Reactivating Solution (see Main-
tenance)
Check connector
(See above)
Rehydrate (see Maintenance)
(See above)
© 04 / 16 Mettler-Toledo GmbH
en
9
53 500 086

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis