Example: Determining the density of a solid sample.
Parameter settings:
Applic.: Density: Dec.Plcs: 1 Dec. Pl. (menu code 2. 9. 1. 2)
Step
1. Attach sample holder to suspension wire
2. Tare the balance
3. Start application program
4. Confirm "Air" display
5. Determine the weight
of the sample in air:
Place sample on the balance
6. Store value for weight in air
7. Remove sample from the balance
8. Determine weight in liquid:
place sample in holder
9. Confirm "water" display
10. Immerse sample in liquid
11. Store value for weight in
liquid, view result, and print
12. Delete result
13. Repeat as desired, starting from Step 3.
Key
(or instruction)
t
ENTER
ENTER
ENTER
ENTER
ENTER
CLEAR
Display/Data output
0.0
g
Air
?
20.0
+
g
*
water
?
0.0
g
*
15.0
+
g
*
o
4.0
+
*
Wa
+
20.0 g
Wfl
+
15.0 g
Rho
4.0 o