Internal Calibration/Adjustment (Only on SI models)
Set the following parameters:
Setup: Bal.Scal.: Cal.-Adj.: Cal.Int. (menu code 1.1. 9. 4)
The built-in motorized calibration weight is applied and removed automatically
for internal calibration.
Step
1. Tare the balance
2. Start calibration
The built-in weight is
applied automatically
3. Calibration/adjustment executed
4. The built-in weight is removed
16
Key (or instruction)
t
q
Display
0.0
g
Cal.Int.
Cal. run.
Cal.end
0.0
g