Herunterladen Inhalt Inhalt Diese Seite drucken

Operation; Environmental Conditions; Troubleshooting - VOLTCRAFT SL-10 Bedienungsanleitung

Schallpegel-messgerät
Vorschau ausblenden Andere Handbücher für SL-10:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation

When taking measurements, always point the measuring microphone (2) exactly
towards the sound source to be measured.
Ensure that there are no persons or objects between the measuring microphone (2)
and the sound source.
To avoid affecting the sound waves, hold the measuring device as far away from
your body as possible or place it on a stand.
Avoid vibrations or movements.
Use the enclosed wind deflector (1) for wind or strong draughts so as not to distort
the measurement through wind sounds. The wind deflector (1) does not distort the
measurement results.
• Press the
button once to switch on the device.
• Press the
button once to switch off the device.
The device turns off automatically after approx. 5 minutes of inactivity.
To deactivate the automatic switch-off for the next measurement, press the
ton with the device switched off and then press and hold the
(3) shows the respective status ("APO OFF" = automatic switch-off deactivated).
The automatic switch-off is activated when the device is switched on the next time.
• When switched on, the sound level meter immediately begins to measure the sound level and
displays the measured values in the measurement value display (8).
• Press the
button to toggle between maximum measurement value, minimum measu-
rement value and current measurement value. The currently displayed values are indicated
with the MAX/MIN display (5) (nothing displayed >> current measurement value).
• Press the
button to "freeze" the displayed value. The current measurement value is
shown continuously, HOLD (6) is shown in the display. Press the
to the normal display.
• Press and hold the
button to switch the display light on or off.
• Press the
button to toggle the measurement interval. FAST/SLOW (9) in the display
indicates the set measurement interval (FAST = 125 ms, SLOW = 1 s).
The FAST setting is used for quickly changing sound levels. The SLOW setting is
used for continuous, constant sound levels.
• If the measured values are outside of the measurement range, the display (3) shows "UN" or
"OL" ("UN" = measurement values < 30 dB, "OL" = measurement values > 130 dB).
Main points
Button
short confirmation
Switching off
Toggling of display type
Toggling of measurement interval
"Freeze" measurement value

Troubleshooting

The device cannot be switched on, the display (3) does not switch on
• Are the batteries spent?
The displayed measurement value is incorrect or changes when the measurement
signal changes
• Are the batteries spent?
• Is the signal to be measured inside of the measurement range?
• Are there obstacles or persons present between the sound source and the measurement
microphone (2)?
• Was the maximum or minimum measurement value display activated?
• Was the measurement value display "frozen"?
button. The display
button again to return
long confirmation
Switching on
Display light on/off
Together with
Deactivating of automatic switch-off
Care and cleaning
Do not use abrasive cleaning agents, cleaning alcohol or other chemical solutions
since these could damage the housing or even impair operation.
• The product is maintenance-free for you, apart from the occasional battery replacement.
Inside the product are no components that need to be maintained by you.
• Use a dry, lint-free cloth for cleaning.
Disposal
a) Product
Electronic devices are recyclable and do not belong in the household waste. Dispo-
se of an unserviceable product in accordance with the relevant statutory regulations.
Remove the batteries that may be inserted and dispose of them separately from the
product.
b) Batteries
As the end user, you are required by law (Battery Ordinance) to return all used batteries/rechar-
geable batteries; disposal in household waste is prohibited!
Batteries/rechargeable batteries contain harmful materials and are labelled with the
symbol shown to indicate that disposal in the household waste is forbidden. The
symbols of the critical heavy metals are: Cd=cadmium, Hg=mercury, Pb=lead (mar-
king can be seen on the battery / rechargeable battery, e.g., underneath the refuse
bin symbol shown on the left).
You can return your used batteries/rechargeable batteries free of charge at the official coll-
ection points of your community, in our stores, or at places where batteries or rechargeable
batteries are sold!
You thereby fulfil your statutory obligations and contribute to the protection of the environment.
Technical data
but-
Operating voltage ......................... 4.5 V/DC
Batteries ....................................... 3 x 1.5 V micro-cell (AAA)
Battery life ..................................... approx. 10 h
Automatic switch-off ..................... approx. 5 min
Measurement range ..................... 30 - 130 dB (A)
Frequency weighting .................... A-weighting (human hearing frequencies)
Frequency range .......................... 31.5 – 8000 Hz
Accuracy ....................................... ±1.5 dB (@1 kHz, standard sound source)
Measurement interval ................... 125/1000 ms
Display .......................................... 4 digits
Display resolution ......................... 0.1 dB
Microphone ................................... 12.7 mm condenser microphone
Dimensions ................................... 155 x 50 x 27 mm
Weight .......................................... 105 g

Environmental conditions

Operating conditions ..................... 0 to +30 °C / max. 80% RH
Storage conditions ........................ -20 to +60 °C / <80% RH
Height .......................................... 0 - 2000 m above sea level
This is a publication by Conrad Electronic SE, Klaus-Conrad-Str. 1, D-92240 Hirschau (www.conrad.com).
All rights including translation reserved. Reproduction by any method, e.g. photocopy, microfilming, or the capture in
electronic data processing systems require the prior written approval by the editor. Reprinting, also in part, is prohibited.
This publication represent the technical status at the time of printing.
© Copyright 2018 by Conrad Electronic SE.
>+30 to +40 °C / max. 80% RH
>+40 to +50 °C / max. 45% RH
1662850_V2_0818_02_m_VTP_en

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Vc-83142501662850

Inhaltsverzeichnis