• Never switch the measuring device on immediately after it has
been brought from a cold into a warm area. The resulting con-
densation water could damage your device. Allow the device
to come to room temperature before switching it on.

4. Operating elements

Figure see swing outside
1
1 3-
/
-digit LCD display (LCD = liquid crystal display)
2
2 "Outer" connector strip(s) with the following pin configurati-
on (Rear view from right to left):
2 -V-range Bridge here for measurements
up to max.
20 --V-range Bridge here for measurements
up to max.
20
200 -V-range Bridge here for measurements
up to max.
200
600 -V-range Bridge here for measurements
up to max.
600
COM = Common
= Measuring input "-", reference ground
N.C.
= Not connected
Vi
= Voltage in
= Measuring input "+"
N.C.
= Not connected
GND = Ground
= Supply input "-" = ground
9 V
= Supply input "+" 9 VDC (stabilised)
3 "Inner pin strip" = JP2 for fixing the decimal point and the unit
of measurement "V" or "mV".
6
2
V
(solder bridge)
V
V
V
5. Installation, settings, commissioning, perfor-
ming measurements
5.1 Settings
Before the panelmeter is installed, the decimal point and the
unit of measurement must be set, depending on the voltage
measuring range. To do this, proceed as follows:
a) Open the panel housing by carefully unscrewing the four
fixing screws with a suitable crosshead screwdriver.
b) Remove the rear housing shell.
c) The measuring board is now exposed. Remove this from the
mounting (pin socket at the front).
d) The jumpers are now accessible. In the following tables the
"pin contact pairs" are shown from left to right.
e) In order to set the decimal point, observe the following
table:
Decimal point position
Jumper JP2
0000
-
000.0
-
00.00
-
0.000
-
Where: "-" = not used; "0" = no jumper; "S" = jumper, i.e. two
pins bridged
f) When the decimal point has been set, the unit of measure-
ment must be set . Note the following table:
-
-
0
0
0
-
-
0
0
S
-
-
0
S
0
-
-
S
0
0
7