16
Terminal compartment B
Signal circuit in type of protection Intrinsic Safety "ia"
Signal circuit
PA, FF *
1
U
 17.5 V
i
+ –
I
 500 mA
i
 5.5 W
P
i
or
U
 24 V
i
I
 250 mA
i
P
 1.2 W
i
C
 5 nF
i
L
 10 μH
i
suitable for connection to a fieldbus system according to the FISCO model
1
*
only available at the versions: FMG60-**J******, FMG60-**K******, FMG60-**L******.
PT100
PT100
U
= 8.4 V
o
I
= 8.3 mA
o
P
= 17.5 mW
o
= 1012 
R
i
C
= 0
i
L
= 0
i
Characteristic curve: linear
Cascade out
U
= 8.4 V
o
– +
I
= 19.2 mA
o
P
= 40.3 mW
o
R
= 439 
i
C
= 5.3 nF
i
L
= 67 μH
i
Characteristic curve: linear
Only for connection to FMG60 signal circuit "Cascade in"
Cascade in
U
= 8.4 V
i
+ –
I
= 19.2 mA
i
P
= 40.3 mW
i
C
= 0
i
L
= 67 μH
i
Only for connection to FMG60 signal circuit "Cascade out"
Connection
U
= 4.7 V
o
for FHX40
I
= 37.7 mA
o
P
= 44.3 mW
o
= 125 
R
i
C
= 12.7 nF
i
L
= 0
i
Characteristic curve: linear
PT100
Ex ia IIIC
C
= 6 μF at
o
L
= 1 mH
o
C
= 5.2 μF at
o
L
= 2 mH
o
C
= 42 μF
o
L
= 199 mH
o
For connection to the approved display FHX40 with
associated cable (KEMA 02 ATEX 1203) in type of
protection Intrinsic Safety Ex ia IIC or IIIC.
Observe associated Safety Instructions XA00193F!
C
= 150 μF
o
L
= 25 mH
o
Gammapilot M
OFF
ON
SIM WP
9 10 1112
IN CASCADE OUT
+ –
+ –
– +
3 4
5 6
7 8
Endress+Hauser

Werbung

loading