4.3 Adjust temperature
The temperature is measured at two selectable points and it is possible to correct it
there.
Invoke service function „Tmp Adj", see chap. 4.1.
Confirm by pressing the PRINT button. The current setting for the first tempera-
ture point is displayed. Or leave it on factory setting 80°C or select a desired tem-
perature with the help of the navigation buttons
Confirm with the PRINT button, the first heat-up phase is started. The remaining
time is displayed.
Temperature calibration of first point will take 15 min. Compare the temperature
displayed on the thermometer with that of the moisture analyzer. If the two values
do not match, correct them using the navigation buttons
PRINT button
The current setting for the second temperature point is displayed. Or leave it on
factory setting 150°C or select the desired temperature with the help of the navi-
gation buttons
the first one, max. 160°C.
Confirm with the PRINT button, the second heat-up phase is started. The remain-
ing time is displayed.
Temperature calibration of the second point will take 15 min. Compare the tem-
perature displayed on the thermometer with that of the moisture analyzer. If the
two values do not match, correct with the help of the navigation buttons
Confirm by pressing the PRINT button. The adjustment is completed, the appli-
ance returns into the menu. Use the ON/OFF button to return into weighing mode.
Due to the AUTO-POWER OFF function the display unit of the temperature
calibration set switches off automatically during adjustment. Turn-on appli-
ance again, the current temperature adjustment will not be disturbed.
DLB_A01-BA-def-1110
. The second temperature point has to be at least 30°C over
(available 50-130°C).