Herunterladen Inhalt Inhalt Diese Seite drucken

KERN DLB_A01 Betriebsanleitung Seite 14

Temperatur-kalibrierset
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.2 Calibrate temperature
During temperature calibration only a check is carried out, i.e. no values are changed.
Invoke service function „Tmp test", see chap. 4.1.
Confirm by pressing the PRINT button. The current setting will be displayed. Use
the navigation buttons
35- 160°C.
Confirm using the PRINT button, the appliance heats up to the set temperature.
After approx. 15 minutes compare the temperature displayed on the thermometer
with that of the moisture analyzer. If the two values do not match, we recommend
a temperature adjustment, see chap. 3.
Use the ON/OFF button to finish the calibration.
Use the ON/OFF button to return into weighing mode.
14
to select the temperature during the check, available
DLB_A01-BA-def-1110

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis