Failure: The reading does not change, when the
electrode is immersed in the aquarium.
Cause: The cap has not been removed from the
electrode.
Remedy: Remove the cap for the measurement.
Cause: The electrode is soiled or too old.
Remedy: Clean the electrode or replace, if and
when necessary.
Failure: The display renders excessively high
values.
Cause: The platinum layer of the electrode is
covered with deposits.
Remedy: Clean the electrode.
Failure: The display renders too low values.
Cause: The electrode is soiled, too old, defective
or dried up.
Remedy: Clean the electrode or replace, if and
when necessary.
mV-Controller Set 7075/2
Disfonctionnements: L'affichage ne varie pas
lorsque l'électrode est plongée dans l'eau.
Raisons: Le capuchon de protection est en bout
d'électrode.
Solutions: Le capuchon doit être retiré pour
effectuer la mesure.
Raisons: L'électrode est sale ou usagée.
Solutions: Nettoyez l'électrode, évent. renouvelez.
Disfonctionnements: Valeur trop élevée sur
l'affichage.
Raisons: Dépôts sur la surface platine de
l'électrode.
Solutions: Nettoyez l'électrode.
Disfonctionnements: Valeur trop faible sur
l'affichage.
Raisons: Electrode sale, trop âgée, défectueuse
ou sèche.
Solutions: Nettoyez l'électrode, évent. renouvelez.