5.3.2 Installation with tensile force adapter and knuckle eye

If the transducer is to be tensile force loaded, it can be mounted with an
adapter (HBM accessory) and a knuckle eye. A centering hole is located on
both sides of the transducer.
When knuckle eyes are used, the breaking force is reduced to 150 %.
Knuckle eyes are only suitable for use with quasi‐static loading (alternating
loads v10 Hz).
Installation with flange through‐hole
M
A
Installation with knuckle eyes
Fig. 5.2
Installation for tensile loading
Nominal force
H1 (mm)
(kN)
100
app.159
200
app. 203
500
app. 319
1)
secured with 2 screws to prevent torsion
A0809-8.0 en/de/fr
provided by customer
Tensile force adapter
lock nut (M
)
B
Knuckle eye ZGUW
Bearing plate
H2 (mm)
Starting torque M
app. 269
app. 351
app. 575
Installation with flange thread
M
A
provided by customer
lock nut (M
)
B
Tensile force adapter
M
A
M
A
lock nut (M
)
B
Knuckle eye ZGUW
Starting torque M
A
(NVm)
115
280
560
13
Tensile force adapter
lock nut (M
)
B
Knuckle eye ZGUW
T
B
(NVm)
(mm)
1900
app. 15
4300
app. 19
1)
-
app. 23