At the completion of a test, information can be entered in the entry boxes (date
of the test, number of the NANOCOLOR
®
USB-T-Set, the certificate number
applicable and the date of issue). In the "Notes box", any notes that may be
relevant can be entered.
Note:
If the heating block does not contain any data or the information has already been
entered, it is not possible to input any data or amend any existing data (the fields
are greyed). The input of data or any modification is possible, only after a new test
run has been completed.
The temperature table is divided into two columns. In the "nominal" column, all tem-
peratures are listed that have been programmed in the heating block. The "actual"
column shows the temperatures last measured with the NANOCOLOR
®
USB-T-
Set.
When the "Report" button is activated, the NANOCOLOR
USB-T-Set software
®
creates a certificate of the test run, that is then automatically printed.

3.6 Calibration

After completing a calibration run, the calibration data measured by the
NANOCOLOR
®
USB-T-Set and saved in the heating block, can be read out,
displayed and edited.
After calibration, information can be entered in the entry boxes (date of the
calibration, number of the NANOCOLOR
USB-T-Set, the certificate number
®
applicable and the date of issue). In the "Notes box", any notes that may be
relevant can be entered.
Note:
If the heating block does not contain any calibration data or the data has already
been entered, it is not possible to input any data or amend any existing data (the
fields are greyed). The input of data or any modification is possible, only after a
new calibration run has been completed.
The temperature table is divided into three columns. In the "nominal" column, all
temperatures are listed that have been programmed in the heating block. The
"actual" column shows the temperatures measured with the NANOCOLOR
®
USB-T-Set. In the "Calibrated to" column, all calibrated temperatures are listed.
When the "Report" button is activated, the NANOCOLOR
®
T-Set software creates
a certificate of the calibration run, that is then automatically printed.
16
NANOCOLOR
®
USB-T-Set / 09.15