Connecting The Voltage Supply - Checkline Ne214 Betriebsanleitung

Elektronischer
Verfügbare Sprachen

Verfügbare Sprachen

NE214
AC voltage connection
115 VAC
230 VAC
(24 VAC)
(48 VAC)
32
Connecting assignment
Terminal
Function
1
Not assigned
2
Voltage supply (-)
3
Voltage supply (+)
4
Sensor supply 0 V
5
Sensor supply +24 V
6
Signal input track A
7
Signal input track B
8
Reset
9
Signal stop
10
Preset 1 (collector)
11
Preset 1 (emitter)
12
Preset 2 (collector)
13
Preset 2 (emitter)
14
Reserved for option RS485 T,R-
15
Reserved for option RS485 T,R+
16-19
Reserved for option RS232 or RS422
20-22
Reserved for option relay output limit value 1
23-25
Reserved for option relay output limit value 2
27
Reserved for option analog output
Reserved for option analog output (jumper to U)
28
29
Reserved for option analog output (I)
30
Reserved for option analog output (U)
For protection against shock hazards as specified in EN 61010, stran-
ded conductors may only be connected using wire end ferrules with
insulating caps. Terminals which are not assigned in the factory must
not be otherwise assigned by the user. We recommend shielding
all encoder connecting leads and earthing the shield at one end.
Earthing at both ends is recommend to avoid RF interference or if
equipotential bonding conductors are installed over long distances.
Sensor connecting leads should not be laid in the same trunking as
the mains power supply cable and output contact leads.

3.1 Connecting the voltage supply

It is possible to switch between two different alternating voltage
ratings by means of the voltage selector switch at the side of the
unit. The higher of the two alternating voltage ratings (48 VAC or
230 VAC) is preset by the factory.
➜ Set the required alternating voltage with the voltage selector
switch.
➜ Connect the supply voltage to connections 2 and 3 in accor-
dance with the terminal diagram.