Chemical cleaning with disinfection
Cleaning and disinfection can be carried out in a single cycle
using a chemical preparation process. For this, use the clean­
ing disinfecting agent Bomix
To ensure safety when handling chemicals, follow the instruc­
tions for use of the disinfecting agent, particularly the accompa­
nying safety instructions.
CAUTION
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection. Adequate cleaning
with disinfection can only be assured if the specified mixing
ratio and application time are adhered to, and if all individual
parts are completely immersed in the solution for the entire
application time. There must not be any cavities or air
bubbles.
• Prepare a 2% solution of
®
Bomix
 plus by mixing 10 ml of
the concentrate with 500 ml
tap water.
• Place all the individual parts in
the prepared solution and
leave them to soak for 5 min.
If the application period is exceeded significantly, the plastic
parts may take on the smell of the disinfectant.
• Rinse off all parts thoroughly in running water (residues of
the disinfectant can cause allergic reactions or irritations of
the mucous membrane).
• You can remove excess water more quickly by shaking all
parts.
• Dispose of the used solution (the diluted solution can be got
rid of down the drain).