Use as a household thermometer (scan mode)
The thermometer can also be used in the house for
non-contact temperature measurements in the range of
–22.0 °C - 80.0 °C (–7.6 °F - 176.0 °F).
Attention! Do not use this mode for measuring
body temperature.
– With this measurement, the current surface temperature
of the object is displayed. It can be diff erent from its
internal temperature, especially if the surface is exposed
to direct sunlight or a draught!
Due to its design, the thermometer measures the precise
value when its casing has adjusted to room temperature.
That is why you should not hold it in the hand for a
long time when you want to take a measurement and
why you should keep it away from direct sunlight!
Do not use the appliance immediately after cleaning
the measuring head
of the detergent may falsify the measuring result.
Sample applications:
-
The temperature of milk in baby bottles
-
The temperature of bathwater
1. If the sleep mode has been activated, activate the
thermometer again by pressing any button before
taking a measurement.
2. Carefully press on the protective cap
it with a gentle rotary motion and without using any
force.
82 GB
. The evaporative cooling eff ect
1
and remove
8