BATTERy chARGING & FILTER mAINTENANcE
Battery charging (7314P Series Only)
On 7314P series, the units are equipped with a factory-installed rechargeable battery. The unit will have a light for low battery and charge indication.
1
2
Ensure power switch is
"off". Plug the small connector of
the AC or DC adapter into the
DC power input.
NOTE– A discharged battery will require 10-17 hours (depending on depth of discharge) of charging to reach a full capacity.
NOTE– Do not connect the AC adapter to an outlet controlled by a switch to ensure power is supplied to unit at all times.
NOTE– Do not connect the DC power cord to an outlet that is not constantly energized.
NOTE– A fully charged battery will provide approximately 60 minutes of continuous operation at a zero vacuum level (free flow). Operation time will decrease with higher
vacuum levels.
NOTE– If unit is not in use for extended periods, battery should be recharged every 6 months minimum.
CAUTION– Discharging the battery completely will shorten the life of the battery. Do not operate the unit more than a few minutes if the low battery indicator light is lit.
Recharge as soon as possible.
NOTE– When charging the battery, use an external power source and verify that the charge light illuminates when the unit is in the "Off' position. If the battery does not
charge, please be sure the model you are using has a battery installed prior to contacting your authorized DeVilbiss Healthcare provider.
NOTE– The internal rechargeable battery is sealed lead-acid. Contact your local authorities for instruction on proper disposal.
changing Filter cartridge (Disposable container)
Change filter cartridge if overflow occurs or every two months, whichever comes first.
1
2
Remove filter cartridge and
Turn unit
"off".
43/8" tubing.
NOTE– Do not substitute any other material for this filter cartridge. Substitution may lead to contamination or poor performance; use only DeVilbiss filter cartridges.
NOTE– The filter cartridge contains a hydrophobic filter. If the filter media becomes wet, air flow will be stopped. The filter cartridge must then be replaced. Do not remove
filter media from filter cartridge.
changing Bacteria Filter (Reusable container)
Change bacteria filter if overflow occurs or every two months, whichever comes first.
1
2
Remove filter by disconnecting
Turn unit
"off".
it from suction unit and lid
assembly.
NOTE– Do not substitute any other material for this bacteria filter. Substitution may lead to contamination or poor performance; use only DeVilbiss filters.
NOTE– Bacteria filter must be changed between patients.
A-7314
3
4
Plug the other end into an AC
Battery charging begins; 10-17
wall outlet or DC receptacle.
hours for full charge.
3
Install new cartridge and
attach tubing.
3
Replace with a clean DeVilbiss
bacteria filter (non-sterile) and
remount to suction unit and lid.
Ensure that the clear side of the
bacteria filter is toward elbow
and bottle when installing. Do
not reverse direction of filter.
Additional filters (7305D-608
12/pack) may be purchased
from your authorized DeVilbiss
Healthcare provider.
5
Battery charging complete.
EN - 77