Herunterladen Inhalt Inhalt Diese Seite drucken

TTI TG4001 Bedienungsanleitung Seite 7

40mhz dds function/arbitrary generator
Inhaltsverzeichnis

Werbung

Sum In
Frequency Range:
Signal Range:
Input Impedance:
Ref Clock In/Out
Set to Input:
Set to Output:
Set to Phase Lock:
Maximum Input Voltage:
INTERFACES
Full remote control facilities are available through the RS232, USB or GPIB interfaces.
RS232:
IEEE−488:
USB
GENERAL
Display:
Data Entry:
Stored Settings:
Size:
Weight:
Power:
Operating Range:
Storage Range:
Environmental:
Options:
Safety:
EMC:
6
DC − 30 MHz .
Approximately 2 Vpk−pk input for 20Vpk−pk output; maximum input
±10V.
Typically 1kΩ.
Input for an external 10MHz reference clock. TTL/CMOS threshold
level.
Buffered version of the internal 10MHz clock. Output levels nominally
1V and 4V from 50Ω.
Used together with SYNC OUT on a master and TRIG IN on a slave to
synchronise (phase lock) two separate generators.
+5V, –1V.
Variable Baud rate, 38400 Baud maximum. 9−pin D−connector.
Optional - Conforms with IEEE488.1 and IEEE488.2
1.1
20 character x 4 row alphanumeric LCD.
Keyboard selection of mode, waveform etc.; value entry direct by
numeric keys or by rotary control.
Up to 9 complete instrument set−ups may be stored and recalled from
non-volatile memory.
3U (130mm) height; 212mm (½−rack) width; 335mm long.
4.1kg (9lb).
220-240V nominal 50/60Hz; 110-120V or 100V nominal 50/60/400Hz;
nominal voltage adjustable internally; operating range ±10% of
nominal; 60VA max. Installation Category II.
+5°C to 40°C, 20−80% RH.
−20°C to + 60°C.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
19 inch rack mounting kit, GPIB remote control interface.
Complies with EN61010−1.
Complies with EN61326