schwa-medico Instruction Manual - AS SUPER 4 digital
Free programs
(own parameter settings)
The frequency, the pulse width and the therapy duration of the programs no.
1.0, 1.1, 1.2, 1.3, 3.0, 3.1 and 3.2 can be modified for own parameter settings.
Setting parameter for program no. 1.0, 1.1, 1.2, 1.3
(standard stimulation)
Select the program with the P and E key.
Press the E key for 4 seconds to enter the parameter modification mode. This is
indicated by an acoustic signal and the word "SET" flashes on the display.
Press ▲▼ key of channel 1 or 3 to modify the intensity between 1 Hz and 120 Hz.
Press ▲▼ key of channel 2 or 4 to modify the pulse width between 50 µs and
300 µs.
Press the E key to save the changes.
Press ▲▼ key of channel 1 or 3 to modify the time between 10 min and 90 min.
4 seconds after pressing the last key, the device automatically returns to the
stimulation mode. This is indicated by an acoustic signal.
The stimulation can then be started by increasing the intensity with the ▲ keys
of each connected channel.
Setting parameter for program no. 3.0, 3.1, 3.2 (HAN stimulation)
Select the program with the P and E key.
Press the E key for 4 seconds to enter the parameter modification mode. This is
indicated by an acoustic signal and the word "SET" flashes on the display.
Press ▲▼ key of channel 1 or 3 to modify the intensity between 1 Hz and 120 Hz.
Press ▲▼ key of channel 2 or 4 to modify the pulse width between 50 µs and
300 µs.
Press the E key to save the changes.
Press ▲▼ key of channel 1 or 3 to modify the time between 10 min and 90 min.
Press the E key to save the settings of the first frequency. You can now set the
second parameters which will alternate every 3 seconds with the first parame-
ters you have selected.
Press ▲▼ key of channel 1 or 3 to modify the intensity between 1 Hz and 120 Hz.
Press ▲▼ key of channel 2 or 4 to modify the pulse width between 50 µs and
300 µs.
4 seconds after pressing the last key, the device automatically returns to the
stimulation mode. This is indicated by an acoustic signal.
The stimulation can then be started by increasing the intensity with the ▲▼
keys of each connected channel.

Factory settings

Press the
key for 4 seconds to retrieve the factory settings and delete the
changed parameters in the AS SUPER 4 digital. This is indicated by an acoustic
signal.