STORAGE AND MAINTENANCE

• Make sure no residual liquid in the nebulizer head
• Make sure the nebulizer head has been cleaned and disinfected thoroughly after each use.
• Store the device and the accessories in clean and safe place.
• Protect the device against direct sunshine and shock.
• Remember to keep any heating device and open fire far away from the storage place.
• Protect the device against contact with any corrosive liquid.
• Never dry the device and its accessories with microwave oven.
• Never wind any power cord around the main unit.
• Always contact your supplier or the nearest local distributor if repair is necessary.
• The warranty period of this device is one (1)year from the date ofpurchase.
• Always use or store the device in such environment that the tempera-ture and humidity follow the
requirements as specified in the section-SPECIFICATIONS,otherwise the device could not function well.
Warning
• Before the device is send back to the factory for inspection and repair, never dismantle or attempt
to repair it without permission.
TROUBLESHOOTING
NO. Faults
1
Lower nebulizing rate
Power indicator
does not light upand
2
the devicedoes not
work
Power indicator lights
3
up but the device
does notwork
The nebulizer auto-
4
matically shuts down
during use
Symptom/Possible Causes
Low battery indicator lights up/
Battery exhausted
The metal mesh holes are clogged
The medication will bedepleted or
not in contact with metal mesh for
morethan 10 seconds
The metal mesh is damaged
Low battery voltage
The indicator does not light after
charging
Incorrect installation of nebulizer
Low battery indicator lights up/
Battery exhausted
Electrodes on the main unitor nebuli-
zer are dirty
The angle of inclination of the nebuli-
zer is too large and the medication is
not incontact with the metal mesh.
The angle of inclination of the nebuli-
zer is too large and the medication is
not in contact with the metal mesh.
Shake the nebulizer vigorously during
use
Solutions
Charge the battery and restart
Clean the nebulizer (please
refer tothe section-CLEANING
AND DISINFECTION); change
nebulizing head if necessary
Tilt the front of the nebulizer
to allow the medication to
contact the metal mesh
Replace the nebulizer
Use Type-C cable to power
the batter for a period and
then use it again
Check whether the charging
cable is well and connected
correctly, and the adapter is
normal
Reinstall the nebulizer
correctly
Charge the battery and restart
Clean the electrode with medi-
cal gauze
Reduce the tilt angle of the
nebulizer or tilt the front of the
nebulizer to the user to allow
the medication to contact the
metal mesh
Reduce the tilt angle of the
nebulizer or tilt the front of the
nebulizer to the user to allow
the medication to contact the
metal mesh
Hold the nebulizer smoothly
when using