8
Troubleshooting
Problem
The viscometer does not work
The instrument does not show "zero" without any
fluids
Viscosity speed is unstable and inaccurate
9
Further information
9.1
Calibration
The viscometer comes factory-calibrated and with a calibration certificate.
We recommend checking the accuracy of the viscometer on a regular basis by using optional calibration
oils.
Model
PCE-RVI 4 VP 20
PCE-RVI 4 VP 60
For the verification, we recommend using the R4 spindle. Perform at least 5 measurements and compare
the reading to the viscosity of the calibration oil.
Note: Keep in mind that the viscosity is greatly influenced by the temperature. Therefore, you should
perform the verification under the specified conditions of the calibration oil.
9.2
Reference substances and viscosities
Substance
Motor oil SAE 10
Olive oil
Coatings (airbrushed)
Yoghurt
Sugar solution 70 %
Lubricating oil
Concentrated juice
Inks
Honey
Nanocellulose (common additive)
Toothpaste
Manual
Possible solution
Check the batteries
Use the auto zero function
Check the rheological characteristics of the fluid
Perform a calibration check by using calibration
standards
Check if the spindle disk rotates steadily
Check if the temperature of the sample is stable
Viscosity standard (suggested)
Approx. 5,000 mPa s
Approx. 2,500 mPa s
Approx. viscosity (mPas)
65
84
100
150
400
50 – 1,000
1,500
550 – 2,200
10,000
8,000 – 10,000
100,000
9