3.3.
Installation and ambient conditions
Set up BINDER Pure Aqua Service upright on an even surface. The connected cartridge is a pressurized
device. The content may irritate eyes and skin.
Risk of injury through opening the cartridge by force or damaging it.
Irritation of skin and eyes. Damage to the cartridge and its environment.
∅ Do NOT open the cartridge.
 Prevent the cartridge from tipping over.
Permissible ambient temperature range during operation: +4 °C / 39.2 °F up to +40 °C / 104°F (climatic
chamber: +18 °C / 64.4°F to +32 °C / 89.6°F.)
Permissible ambient humidity: 85 % r.h. max., non-condensing.
To avoid any possible water damage, provide a floor drain at the location of the device. Select a suitable
installation site to avoid any consequential damage by splashing water.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the unit
in a way that the power supply unit is easily accessible and can be easily disconnected in case of danger.
The unit must not be installed and operated in potentially explosive areas.
Explosion hazard through explosive dust or air-solvent mixture in the vicinity of the
device
Serious injury or death from burns and / or explosion pressure.
∅ Do NOT operate the unit in potentially explosive areas.
 KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.