6.2

Disinfecting

6.3

Sterilizing

7

FUNCTION CHECK

1. Disinfect the device at regular intervals using a disin-
fectant suitable for sensitive surfaces and acrylic glass.
seca recommends a 70 % ethanol solution as a sur-
face disinfectant.
2. Follow the instructions for use of the disinfectant.
3. Disinfect the device as follows:
► Moisten a soft cloth with disinfectant and wipe
down the device with it.
► Comply with the intervals, see table:
Interval
and
Before
after
ment with direct skin contact
As required
The device must not be sterilized.
► Perform a function check before each use.
A complete function check includes:
• A visual inspection for mechanical damage
• A test of the alignment of the device
• A visual and function check of the display elements
• A function check of all the controls shown in the sec-
tion entitled "Overview"
• A function check of the optional accessories
If you find faults or deviations during the function check,
first try to remedy the fault with the aid of the "Trou-
bleshooting" section in this document.
every measure-
Function check • 55