4.4

Establishing power supply

Inserting batteries

Connecting the
power supply unit
(optional)
48 • seca 899
You need 6 batteries (type AA/1.5 V, included in scope of
delivery).
1. Press on the latch of the battery compartment.
2. Remove the cover of the battery compartment.
NOTICE!
Damage to device
Device may overheat and become damaged.
► Ensure the correct polarity of the batteries
(markings in the battery holder).
3. Insert the batteries.
NOTE
If
appears in the display, you have inserted
bAtt
one of the batteries the wrong way round or the
batteries have discharged.
4. Close the battery compartment.
The scale can be operated with a power supply unit (op-
tional accessory).
WARNING!
Personal injury or damage to the device as a result
of incorrect power supply units
Conventional power supply units may deliver a
higher voltage than is indicated on them. The mea-
suring device may overheat, catch fire, melt or
short-circuit.
► Use only original seca power supply units with a
controlled 12 V output voltage.
1. Insert the connector required for your power supply
into the power supply unit.
2. Insert the device connector of the power supply unit
into the connecting socket of the scale.
3. Plug the power supply unit into a power supply
socket.