EN-32
5.2 Calibration
Adventurer Balances (InCal models) offer a choice of six calibration methods: Internal Calibration,
Automatic Calibration, AutoCal™ Adjustment, Span Calibration, Linearity Calibration and Cal Test
(Calibration Test).
Note: The calibration unit is always in grams.
Attention: Do not disturb the balance during any calibration.
5.2.1
Calibration sub-menu (InCal models)
Internal
Calibration
Note: /E models only have Span Calibration, Linearity Calibration and Cal Test.
5.2.2
Internal Calibration
Calibration is accomplished with the internal calibration mass. Internal Calibration can be performed at
any time, provided the balance has warmed up to operating temperature and is level.
With the Balance turned ON and no load on the pan, touch Internal Calibration. Then touch On or Off to
turn it on or off.
Note: When touch Off, press the CAL button on the control panel will start span calibration.
To start calibration, touch Calibration after On is selected.
Note: The default is On.
The Balance begins to calibrate. Follow the onscreen instruction to clear the pan and wait for the process
to finish.
The display shows the status, and returns to the current application after the calibration is finished.
To cancel at any time, press Save.
5.2.3
Automatic Calibration
When Automatic Calibration is set ON, the balance performs a self-calibration:
AutoCal will automatically calibrate the Balance (using the internal mass) each time there is a change in
temperature significant enough to affect accuracy or every 11 hours.
An information screen will appear when an Automatic Calibration is about to start. Three option buttons
will be displayed:
Now
– Press to perform the calibration directly.
5 min
– Press to perform the calibration after 5 minutes.
Deactivate – Press to deactivate the Automatic Calibration function.
Find Quality Products Online at:
Automatic
AutoCal™
Calibration
Adjustment
•
when it senses a temperature change of 1.5°C
•
or every 11 hours
Span
Calibration
Valuetesters.com
ADVENTURER BALANCES
Linearity
Cal
Calibration
Test