1.
Connecting the Weighing Platform to Equipment in a Hazardous Area
Indicator
(e.g., CIS.. with Option Y2)
2.
Connecting the Weighing Platform to Equipment Outside a
Hazardous Area
Non-hazardous area
This equipment may also be
installed in Zone 2 or 22 if it is
suitable for Category 3 according to
the ATEX Directive.
External indicator
or PC, etc.
2)
1:
Equipotential bonding busbar
2:
Connecting cable, cross section (gauge) at least 4 mm
3:
Please follow the (Ex) installation instructions for this equipment
4:
The temperature range may be restricted for operational reasons (please see Data Sheet for platform) .
Date
11.06.04
Written by
11.06.04
Reviewed by
11.06.04
Released by
12
Zone 2 or Zone 22 hazardous area
Gas:
Group II, temperature class T6 (all gases)
Dust: Surface temperature of the equipment 80°C max.
Ambient temperature: -20°C ... +40°C
3
Shielded cable [w/
screen]
for deflecting
electrostatic charges
2)
1)
Nominal power supply
voltage: <= 15Vdc
Name
Designation
Klausgrete
Drawing No.
Weitemeier
Klausgrete
4)
IU.4-......-..... with Option Y2
IF.4-......-.....with Option Y2
Zone 2 or Zone 22 hazardous area
Gas:
Group II, temperature T6 (all gases)
Dust: Surface temperature of the equipment 80°C max.
Ambient temperature: -20°C ... +40°
IU.4-......-..... with Option Y2
IF.4-......-.....with Option Y2
2)
1)
2
Installation Instructions
36116-751-61-A4
or
Alternative 2
4)
Shielded cable (with
shield) for deflecting
electrostatic charges
or
Revision
1
3
Page
of
00