ACL
10
Auto calibration
The WE2111 proceeds ca. every second an
auto-calibration routine (ca. 200ms) in standstill and ca.
every 10 seconds in motion. During that procedure
measurement is impossible. The WE2111 attempts to fill
in the missing readings by sending the last readings
available. At fast processes it might be of advantage to
switch auto-calibration off.
Important
ACL only valid in INDUSTRIAL mode.
Power loss resets the WE2111 i.e. ACL1,1 is valid.
Without auto­calibration the accuracy of the instrument
will drop over longer periods.
General
No. of parameters
Save changes
Increment Trade Counter
Response Time of command
Response Time of query
Parameter Details
Parameter
Description
1
auto-calibration at stand-
still
2
auto-calibration at motion
WE2111, Command Descriptions
2
not possible
no
Range
Default
0,1
1
0,1
1