Calibration
The thermometer is initially calibrated at the time of manufacture. If
the thermometer is used according to the use instruction, periodic
recaliboration is not required. However, We recommend checking
calibration every two years or whenever clinical accuracy of the
thermometer is in question, Please send the complete device to the
dealer or manufacturer. T he above recommendations do not supersede the legal
requirements. The user must always comply with legal requirements for the
control of the measurement, functionality, and accuracy of the device which are
required by the scope of relevant laws, directives or ordinances where the device
is used.A clinical summary and procedures for checking calibration are available
upon request. (Turn on the thermometer and press the button long time until
entering into calibrate mode, software version will be displayed.)
Battery Replacement
1. Replace battery when "
" appears at the buttom of LCD display. (See
Figure 9)
2. Slide battery cover down as shown in Figure 10.
3. Remove battery and install 2 new AA alkaline batteries as shown in
Figure 11 .
4. Slide battery cover back on.
Figure 9
12
Figure 10
Figure 11
Specifications
Product name
Infrared Ear Thermometer
Model
DET-1015
Measuring range
Ear mode: 32.0°C~43.0°C(89.6°F~109.4°F) O
Measuring site
Ear canal(Ear Mode)
Reference body site
Oral (This thermometer converts the ear temperature to
display its "oral equivalent.")
Operation mode
Ear mode(Adjust mode)
Ear mode:±0.2°C (0.4°F) during 35.5°C~42.0°C
(95.9°F~107.6°F)at 15°C~35°C (59.0°F~95.0°F) operating
Laboratory accuracy
temperature range±0.3°C (0.5°F) for other measuring
and operating temperature range
Display resolution
0.1°C or 0.1°F
Measure time
Approximately one second
Ear mode:10°C~40°C(50°F~104°F), 15%~85%RH, non-
Operating
condensing Atmospheric Pressure : 70kPa~ 106kPa
temperature range:
Storage and transport
-25°C~ 55°C (-13°F~131°F), 15%~95%RH, non-condensing
temperature range
Atmospheric Pressure : 70kPa~ 106kPa
Clinical bias:0.08°C( 0.14°F) Clinical repeatability:0.13°C(
Clinical accuracy
0.23°F) Limits of agreement:0.73°C( 1.31°F)
Shock
withstands drop of 3 feet
Dimension
146.4*37.6*57mm
Weight
Approx.117grams(with batteries)
Battery
DC3V(2×AA battery)
Battery life
Approx.6000 readings
Expected service life
Three years
Ingress protecting
IP22
rating
Contraindication
No contraindication
EN
13