storage detects and marks these false echoes, so
that they are no longer taken into account for the
level measurement. A false echo memory should
be created with empty vessel so that all potential
interfering reflections will be detected.
Gating out of false signals
Change now?
Proceed as follows:
1
Move from the measured value display to the
main menu by pushing [OK].
2
Select the menu item "Service" with [->] and
confirm with [OK]. Now the menu item "False
signal suppression" is displayed.
3
Confirm "False signal suppression - Change
now" with [OK] and select in the below menu
"Create new". Enter the actual distance from
the sensor to the product surface. All false
signals in this area are detected by the sensor
and saved after confirming with [OK].
Note:
Check the distance to the product surface,
because if an incorrect (too large) value is
entered, the existing level will be saved as
false signal. The filling level would then no
longer be detectable in this area.
Copy sensor data
This function enables reading out parameter
adjustment data as well as writing parameter
adjustment data into the sensor via the indicating
and adjustment module. A description of the
function is available in the operating instructions
manual "Indicating and adjustment module".
Copy sensor data
Copy sensor data?
Sensor-specific basic adjustment.
1)
Depending on the senso type, see "Technical data".
2)
Special parameters are parameters which are set customer-specifically on the service level with the adjustment
3)
software PACTware™.
VEGASON S 62
Set up with the indicating and adjustment module
Reset
Basic adjustment
If the "Reset" is carried out, the sensor resets the
values of the following functions to the reset values
(see chart):
1)
Function
Max. adjustment
Min. adjustment
Medium
Vessel form
Damping
Linearization
Sensor-TAG
Displayed value
Current output - characteristics
Current output - max. current
Current output - min. current
Current output - failure
Unit of measurement
The values of the following functions are not reset
to the reset values (see chart) with "Reset":
Function
Lighting
Language
HART mode

Factory setting

Like basic setting, in addition special parameters
are reset to default values.
Pointer
The min. and max. distance and temperature
values are reset to the actual value.

Optional settings

Additional adjustment and diagnosis options such
as e.g. scaling, simulation or trend curve presen-
tation are shown in the following menu schematic.
Reset value
0 m(d)
Meas. range end in m(d)
2)
Liquid
not known
0 s
linear
Sensor
Distance
4 ... 20 mA
20 mA
4 mA
<3.6 mA
m(d)
Reset value
no reset
no reset
no reset
3)
31