For your protection
Read the operating instructions in full before st-
arting up and follow the safety instructions.
•
Keep the operating instructions in a place where they can be
accessed by everyone.
•
Ensure that only trained staff work with the appliance.
•
Follow the safety instructions, guidelines, occupational health
and safety and accident prevention regulations.
Work with the device
Before use, check the outlet pres-
DANGER
sure on the pressure reducer and
compare it with the pressure dis-
played on the pressure gauge.
Oxygen as a compressed gas is
oxidising; intensively aids combus-
tion; can react violently with flam-
mable materials.
•
The filling time should not be set to less than 30 seconds.
This will prevent combustible substances from being whirled
around in the crucible.
Use
The IKA
C 48 oxygen filling station is only suitable for filling de-
®
composition vessels C 7010, C 7012 and AOD 1.1 with oxygen.
Unpacking
•
Please unpack the device carefully.
•
In the case of any damage a detailed report must be sent
immediately (post, rail or forwarder).
8
Safety instructions
Correct use and description of the device
Unpacking
•
Insufficient ventilation can cause formation of explosive mix-
tures. Therefore, only use the device in well-ventilated areas.
•
Check the seals for damage before each use.
•
After finishing work, the oxygen supply on the pressure redu-
cer must be closed.
For protection of the device
Maintenance tasks should be per-
DANGER
formed without any excess pres-
sure on the system.
The Valve needle (29) is held in
place with safety varnish to keep it
from being turned unintentionally.
It should only be removed if there
iv no excess pressure on the sys-
tem
•
Keep lines and fittings through which oxygen flows free of
grease.
Description of the device
•
The 2 m pressure hose, included with delivery, can easily be
laid to connect to the oxygen bottle.
•
Therefore, the location of the C 48 filling station does not
depend on the oxygen bottle being in the immediate vicinity.
•
The filling pressure that is set and the pressure in the decom-
position vessel can be read directly from the work station,
thus preventing insufficient filling quantities because of not
observing the minimum filling period.
•
The filling time can be adjusted in increments.
•
There is no need to bee constantly occupied with the oxygen
bottle because the automatic non-return valve only opens if a
decomposition vessel is being filled.
•
A carrier plate (screws on left-handed or right-handed) is
used to prepare samples.
Delivery scope
•
C 48 oxygen filling station
•
Holding plate
•
O-rings
•
Pressure hose (2 m)
•
Operating instruction