The C 48 oxygen filling station requires no maintenance to opera-
te. It is subject only to the natural aging of the components and
their statitical failure rate.
When ordering spare parts, please indicate the manufacturing
number on the rating plate, the device type and the item number
and description of the spare parts.
Please send in only units that have been cleaned and that are free
of hazardous substances for repairs. Use only cleaning agents
approved by IKA
to clean IKA
®
below, use the corresponding cleaning agents:
Dye, coloring
Isopropanol
Construction materials
Water containing detergent
Cosmetics
Water containing detergent/isopropanol
Foodstuffs
Water containing detergent
Combustible materials
Water containing detergent
For materials that are not listed above, please inquire with IKA
Wear protective gloves to clean the units. Electrical devices must
not be placed in the cleaning agent to clean them. Before using
another cleaning or decontamination method than the one re-
commended by the manufacturer, the user must make certain by
consulting with the manufacturer that the intended method will
not damage or destroy the unit.
•
C 29 Pressure gauge, oxygen
•
Holding plate: for inserting decomposition vessels C 7010, C
7011, C 7012, C 7010 Fz C 7012 FZ and AOD 1.1
Permissible operating pressure
Pressure gauge
Typical filling pressure
Pressure hose
Filling speed
Current filling pressure
Oxygen valve
Dimensions (B x T x H)
Weight
Subject to technical changes!
10
Maintenance and cleaning
devices. To clean the residues
®
Accessories
Technical data
bar
bar
m
sec
mm
kg
It should be expected that frequent use will result in wear on the
O rings in the filling head.
The filling head is screwed in place and can be unscrewed after
the pressure hose is removed. The lower O ring (21), which is sub-
jet to particularly high wear and tear, can be reached without di-
sassembling the pressure hose after the cap screw (34) has been
removed.
Item 23 Seal for the non-return
valve
Item 21 Seal for the decomposition
vessel
Repair
In case of repair the device has to be cleaned and free
.
®
from any materials which may constitute a health hazard.
For this you should use the downloaded print of the „Certificate
of compliance" form from the IKA
If you require servicing, return the appliance in its original pa-
ckaging. Storage packaging is not sufficient. Please also use suita-
ble transport packaging.
max. 40
0-50
30
steplessly adjustable
2
> 30
(preset to 30 bar)
readable
automatic non-return valve
200 x 260 x 450
6
34
23
34
21
website www.ika.com.
®