Troubleshooting
If a fault condition should occur, the display on the front panel
Note
will display an error code and an error beep will sound. The
interpretation of the error code, as well as possible causes and
solutions are listed in Table 1.
Note
To recover system from Errors P1-P4, E1-E9, press any button or
re-connect the Probe to the Generator. If errors persist, refer to
Solution section in Table 1. If solutions do not prevent error
occurrence, refer to the Service and Claims section of this manual
to obtain repair service.
Error Code
P1
Probe: Probe not ready
P2
Probe: Probe expired
P3
Probe: Probe identification invalid
P4
Probe: Probe communication error
P5
Probe: Maximum time of use exceeded
E1
Unable to identify error
E2
RF Power too high
E3
RF Voltage too high
E4
RF Current too high
E5
RF delivery exceed countinous use limit • Clear error and continue
E6
Generator temperature too high
E7
Low impedance detected
Switch pressed upon Probe/Footswitch
E8
correction
EN-22
Recoverable Errors & Corresponding Codes
Reason for Error
Solutions
• Check Probe connection to
Generator. Reconnect Probe.
• Discard Probe and replace with
new Probe
• Replace Probe
• Check Probe connection to
Generator. Reconnect Probe.
• Replace Probe
• Discard probe and replace with
new probe
• Check Probe for damage
• Replace Probe
• Check Probe for damage
• Replace Probe
• Check Probe for damage
• Replace Probe
• Check Probe for damage
• Replace Probe
• Ensure proper airflow on sides
and rear of Generator
• Check Probe for damage
• Replace Probe
• Check that switches on Probe,
Footswitch, and Generator are
not pressed