Setting Up the Generator
Initial Set-up of the Generator
Warning Prior to initial use, ensure that all package inserts, warnings,
cautions, and Instructions for Use are read and understood.
Warning Before each use, check that all Generator indicator lights and
audio signals are functional. Make sure that the power cable
plug is properly connected to the Generator receptacle.
1.
The SERFAS Energy Generator should be placed on a Stryker cart or on
any sturdy table or platform. Refer to hospital procedures or local codes
for detailed information.
2.
Provide at least four to six inches of space around the sides of the
SERFAS Energy Generator for convection cooling. It is normal for the
top and rear panel to be warm when system is used continuously for
extended periods of time.
3.
Power cords use with the SERFAS Energy System should comply with
appropriate electrical standards and be suitable for hospital use. Plug the
SERFAS Energy Generator into a grounded receptacle. DO NOT use
extension cords or three-prong to two-prong adapters. The mains
power cord assembly should be checked periodically for damaged
insulation or connectors.
4.
Plug the power cord into back of Generator.
5.
Plug the power cord into AC mains.
EN-11