Selection field
Pumping speed for N
2
Gas throughput at final rotation speed for
Ar
Gas throughput at final rotation speed for
N
2
Fore-vacuum max. for Ar
Fore-vacuum max. for H
2
Fore-vacuum max. for He
Fore-vacuum max. for N
2
Compression ratio for Ar
Compression ratio for CF
4
Compression ratio for H
2
Compression ratio for He
Compression ratio for N
2
Performance curve in gas mode 0, vertex
C
Performance curve in gas mode 0, vertex
D
Performance curve in gas mode 1, vertex
A
Performance curve in gas mode 1, vertex
B
Performance curve in gas mode 2, vertex
E
Performance curve in gas mode 2, vertex
F
Rotation speed variable
Run-up time
Integral leakage rate
Type of electronic drive unit installation
Electronic drive unit
Input voltage(s)
Input voltage: tolerance
Power consumption max.
Current, max.
Permissible irradiated thermal output max.
Sound pressure level
Venting connection
Max. connection pressure (abs.) for vent-
ing/sealing gas valve
Cooling method
Cooling water connection
Cooling water flow
Cooling water temperature
I/O interfaces
Mounting orientation
Permissible radial magnetic field max.
Protection degree
Relative humidity of air
Technical data and dimensions
HiPace® 2300 IU with TC
1200, DN 250 ISO-F
1900 l/s
14 hPa·l/s
40 hPa·l/s
2.4 hPa
1 hPa
1.4 hPa
1.8 hPa
1 · 10
8
1 · 10
8
4
2 · 10
5
3 · 10
1 · 10
8
440/31500 W/min
-1
500/25200 W/min
-1
510/31500 W/min
-1
-1
580/25200 W/min
510/31500 W/min
-1
580/25200 W/min
-1
50 – 100 %
4 min
-8
1 · 10
Pa m³/s
Integrated electronic drive
unit
TC 1200
100 – 120 / 200 – 240 V
AC (±10 %), 50/60 Hz
±10 %
1350 VA
10 A
24 W
≤50 dB(A)
G 1/8"
1500 hPa
Water
G 1/4"
100 l/h
15 – 35 °C
E74, RS-485
90° – 180° (HV flange bot-
tom)
7 mT
IP20, Type 1
5 – 85 %, non-condensing
HiPace® 2300 IU with TC
1200, DN 250 ISO-K
1900 l/s
14 hPa·l/s
40 hPa·l/s
2.4 hPa
1 hPa
1.4 hPa
1.8 hPa
1 · 10
8
1 · 10
8
4
2 · 10
5
3 · 10
1 · 10
8
440/31500 W/min
-1
500/25200 W/min
-1
510/31500 W/min
-1
-1
580/25200 W/min
510/31500 W/min
-1
580/25200 W/min
-1
50 – 100 %
4 min
-8
1 · 10
Pa m³/s
Integrated electronic drive
unit
TC 1200
100 – 120 / 200 – 240 V
AC (±10 %), 50/60 Hz
±10 %
1350 VA
10 A
24 W
≤50 dB(A)
G 1/8"
1500 hPa
Water
G 1/4"
100 l/h
15 – 35 °C
E74, RS485
90° – 180° (HV flange bot-
tom)
7 mT
IP20, Type 1
5 – 85 %, non-condensing