Installation
Fig. 22:
Procedure
1. Use a suitable grounding cable to divert applicative interferences.
2. Route the connection in accordance with locally applicable provisions.
3. Use the turbopump ground terminal (M4 female thread).
5.7.2 Establishing the electric connection
Risk of fatal injury due to electric shock on account of incorrect installation
The device's power supply uses life-threatening voltages. Unsafe or improper installation can lead to
life-threatening situations from electric shocks obtained from working with or on the unit.
► Ensure safe integration into an emergency off safety circuit.
► Do not carry out your own conversions or modifications on the unit.
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installa-
tion leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
► Only connect mating plugs after the mechanical installation.
► Only switch on the vacuum pump immediately prior to operation.
2
1
Tbl. 12:
Preparing a supply cable for the turbopump
There are 2 options for configuring the mains connection for the turbopump.
► Make sure that you have the correct supply voltage.
► Order a corresponding mains connection cable from the Pfeiffer Vacuum accessories range.
► Assemble your own mains connection cable using the HAN 3A connecting socket from the ship-
110/146
Example: Connecting grounding cable
WARNING
WARNING
3
PE
Terminal lay-out of the power supply connector
ment.
M4
Pin
Assignment
1
Phase L
2
Neutral conductor
3
not connected
PE
Earthed conductor