10.1 Technical data

Specification
Product class according to 93/42/EEC
Dimensions W x H x D in cm
Weight in kg
Temperature range
Operation
Storage
Permissible humidity during operation and
storage
Air pressure range
Power supply
Electrical power
Outputs 1-8
Output voltage
Resolution
Socket assignment
Insulation against supply voltage
Insulation against communication signals
Insulation between the 8 channels
Classification as per DIN EN 60601-11:
Type of protection against elec. shock
Protection against harmful ingress of
water and foreign bodies
Classification as per DIN EN 60601-1:
Operating mode
Electromagnetic compatibility (EMC) as
per EN 60601-1-2
Radio interference suppression
Radio interference immunity
+5°C to +40°C
-25°C to +70°C
Rel. humidity 15% to 93%, non-condensing
700 hPa to 1060 hPa, corresponds to a height of approx.
3000 m above sea level
1 channel per 3.5 mm jack socket, tip: signal, sheath: ground
Protection class II
Continuous operation
Test parameters and limit values can be requested from the
manufacturer if required.
IEC 61000-4 Parts 2 to 6, Part 11, Part 8
IEC 61000-3 Parts 2 and 3
10 Appendix
prismaPSG
I
15 x 3 x 9.2
0.2
12 V DC
0.6 VA*
0-1 V
0.002 V
4 kV
2.5 kV
None
IP21
EN55011 B
prismaPSG
EN
39