Herunterladen Diese Seite drucken

VOLTCRAFT PTM-200 Bedienungsanleitung Seite 5

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

8.2 Controlling the display backlight
1. Press the light control button to switch the display backlight on and off.
8.3 Measuring temperature
Preconditions:
a
The temperature of the object, liquid or gas you intend to measure is between -200 to
+200 °C.
1. Insert the measuring probe into liquids, touch objects, or expose the probe to gases to
measure the temperature.
2. Wait for the temperature reading on the display to stabilise. For temperatures below
-50 °C, match the temperature reading against the correction table in the section "Tech-
nical data".
3. Continue taking measurements.
4. Repeatedly press the display switch button MAX MIN to switch between the recorded
maximum reading (display indicator MAX), the recorded minimum reading (display indic-
ator MIN) and the current reading.
9 Maintenance
9.1 Cleaning and care
Important:
Replace the batteries at least once a year to prevent leaking.
Important:
– Do not use aggressive cleaning agents, rubbing alcohol or other chemical solutions.
They damage the housing and can cause the product to malfunction.
– Do not immerse the product in water.
1. Clean the product with a dry, fibre-free cloth.
9.2 Calibration
The thermometer must be calibrated regularly for accurate measurements.
Calibration intervals
Regular calibration interval
Ad-hoc calibration
Calibration service provider
Provider name
Provider contact
Calibration standard
10 Disposal
10.1 Product
This symbol must appear on any electrical and electronic equipment placed on
the EU market. This symbol indicates that this device should not be disposed of
as unsorted municipal waste at the end of its service life.
Owners of WEEE (Waste from Electrical and Electronic Equipment) shall dispose
of it separately from unsorted municipal waste. Spent batteries and accumulators,
which are not enclosed by the WEEE, as well as lamps that can be removed from
the WEEE in a non-destructive manner, must be removed by end users from the
WEEE in a non-destructive manner before it is handed over to a collection point.
Distributors of electrical and electronic equipment are legally obliged to provide free take-
back of waste. Conrad provides the following return options free of charge (more details on
our website):
in our Conrad offices
at the Conrad collection points
at the collection points of public waste management authorities or the collection points
set up by manufacturers or distributors within the meaning of the ElektroG
End users are responsible for deleting personal data from the WEEE to be disposed of.
It should be noted that different obligations about the return or recycling of WEEE may apply
in countries outside of Germany.
10.2 (Rechargeable) batteries
Remove batteries/rechargeable batteries, if any, and dispose of them separately from the
product. According to the Battery Directive, end users are legally obliged to return all spent
batteries/rechargeable batteries; they must not be disposed of in the normal household
waste.
Batteries/rechargeable batteries containing hazardous substances are labelled
with this symbol to indicate that disposal in household waste is forbidden. The ab-
breviations for heavy metals in batteries are: Cd = Cadmium, Hg = Mercury,
Pb = Lead (name on (rechargeable) batteries, e.g. below the trash icon on the
left).
5
Condition
1x per year
If user doubts accuracy
Conrad Electronic SE.
www.conrad.de
www.conrad.com
FFR9903-2014 CM
Used (rechargeable) batteries can be returned to collection points in your municipality, our
stores or wherever (rechargeable) batteries are sold. You thus fulfil your statutory obligations
and contribute to environmental protection.
Batteries/rechargeable batteries that are disposed of should be protected against short cir-
cuit and their exposed terminals should be covered completely with insulating tape before
disposal. Even empty batteries/rechargeable batteries can contain residual energy that may
cause them to swell, burst, catch fire or explode in the event of a short circuit.
11 Technical data
11.1 Temperature correction table
Display (°C)
Actual temperature (°C)
-201.1
-200
-195.5
-195
-190.5
-190
-185.4
-185
-180.4
-180
-175.3
-175
-169.8
-170
-164.7
-165
-159.7
-160
-154.7
-155
-149.6
-150
-144.6
-145
-139.6
-140
-134.5
-135
-129.5
-130
-124.5
-125
11.2 Main unit
Battery
Battery life (approx.)
Ingress protection
Sensor type
Measuring range
Resolution
Sampling rate
Accuracy (-50 to 0 °C) at 25 °C
Accuracy (0 to +100 °C) at 25 °C
Accuracy (+100 to +200 °C) at 25 °C
Calibration standard
Operating temperature
Storage temperature
Operating humidity (non-condensing)
Storage humidity (non-condensing)
Dimensions (W x H x D)
Weight (approx.)
11.3 Measuring probe
Operating temperature (probe handle)
Operating temperature (probe)
Ingress protection
Probe diameter (ø)
Probe length
This is a publication by Conrad Electronic SE, Klaus-Conrad-Str. 1, D-92240 Hirschau (www.conrad.com).
All rights including translation reserved. Reproduction by any method (e.g. photocopying, microfilming or the capture in
electronic data processing systems) requires prior written approval from the editor. Reprinting, also in part, is prohibited.
This publication reflects the technical status at the time of printing.
Copyright by Conrad Electronic SE.
*2584006 _V2_0723_jh_mh_en 27021598595420299-2 I4/O2 en
Display (°C)
Actual temperature (°C)
-119.4
-120
-114.4
-115
-109.4
-110
-104.3
-105
-99.3
-100
-94.2
-95
-89.2
-90
-84.1
-85
-79
-80
-74.5
-75
-69.7
-70
-65
-65
-59.9
-60
-54.9
-55
-49.8
-50
Unit
Value
9 V
h
58
IP65 (front panel only)
PT-1000
°C
-200 to +200
°C
0.1
s
<30
±0.1 % of F.S. ±3 digit
±0.1 % of F.S. ±2 digit
±0.1 % of F.S. ±3 digit
FFR9903-2014 CM
°C
-30 to +45
°C
-20 to +70
%
<80
%
<80
mm
121 x 73 x 30
g
208.5
Unit
Value
°C
-30 to +100
°C
-200 to +200
IP65
mm
3.5
mm
148

Werbung

loading

Diese Anleitung auch für:

2584006