Precision
Shaking Water Baths
Manual Part Number U01366, August 01, 2016
Multilingual Quick Start Guides
Multilingual Essential Safety Instructions
Installation
Operation
Preventive Maintenance
Troubleshooting

Werbung

loading

Inhaltszusammenfassung für Thermo Scientific TSSWB15

  • Seite 1 Precision Shaking Water Baths Manual Part Number U01366, August 01, 2016 Multilingual Quick Start Guides Multilingual Essential Safety Instructions Installation Operation Preventive Maintenance Troubleshooting...
  • Seite 2 Viewing/Changing the Shaker Speed ..................4-3 Changing the Set Point ......................4-3 Selecting a Preset Set Point ....................4-3 Changing a Setting ......................4-4 Selecting a Preset ........................ 4-6 Changing a Preset ....................... 4-6 Temperature Display ......................4-7 Shut Down ........................... 4-7 Thermo Scientific Precision...
  • Seite 3 Chapter 5 Preventive Maintenance ....................5-1 Cleaning ..........................5-1 Electrical Power Cord ......................5-1 Optical Disc ......................... 5-2 Shaker Assembly ......................... 5-2 Chapter 6 Troubleshooting ........................ 6-1 Error Displays ........................6-1 Checklist ..........................6-2 Statement of Conformity Warranty Precision Thermo Scientific...
  • Seite 4 Precision Water Bath Quick Start Guide This quick start guide is intended for initial start up only. For all other procedures you must refer to the manual. Also, if any of these steps are not clear refer to the manual before proceeding. Filling (Only approved fluids are Filtered/Single Distilled Water and De-ionized Water) Ensure the reservoir drain port is closed and that all plumbing connections are securely plumbed or capped.
  • Seite 5 Viewing/Changing the Shaker Speed You can view/change the speed value with the bath running or not. From the home screen press the soft key below to display: Press, and hold, the arrow keys to bring up the desired value. The values are 0 and 30 to 200.
  • Seite 6 Kurzanleitung zur Inbetriebnahme des Präzisionswasserbades Diese Kurzanleitung ist nur für die erste Inbetriebnahme vorgesehen. Für alle anderen Verfahren müssen Sie im Handbuch nachsehen. Auch wenn irgendwelche Schritte unklar sind, schaue Sie im Handbuch nach, bevor Sie fortfahren. Befüllung Stellen Sie sicher, dass der Ablauf des Behälters geschlossen ist und dass alle Anschlüsse sicher angeschlossen oder verschlossen sind. Achten Sie außerdem darauf, dass vor dem Befüllen des Behälters alle Rückstände gründlich entfernt wurden.
  • Seite 7 Anzeigen/ändern der Rüttelgeschwindigkeit Die Geschwindigkeitseinstellung lässt sich sowohl bei eingeschaltetem als auch bei ausgeschaltetem Bad anzeigen/ändern. Vom Einstiegsbildschirm aus drücken Sie die Schaltfläche unten zur Anzeige von: Drücken Sie die Pfeiltasten, und halten Sie diese gedrückt, bis der gewünschte Wert erscheint.
  • Seite 8 Guide de démarrage rapide - Bain de précision Ce guide de démarrage rapide est destiné à la mise en marche initiale uniquement. Pour toute autre procédure, merci de vous référer au manuel. De plus, si l’une de ses étapes ne vous paraît pas claire, téléchargez le manuel avant de commencer.
  • Seite 9 Affichage/modification de la vitesse d’agitation Vous avez la possibilité d’afficher/de modifier la vitesse, que le bain soit en cours d’utilisation ou non. À l’écran d’accueil, appuyez sur la touche se trouvant sous pour afficher : Appuyez sur les touches flèche et maintenez-les enfoncées jusqu’à atteindre la valeur désirée.
  • Seite 10 Guía de inicio rápido de bañeras de agua de precisión Esta guía de inicio rápido se ha creado únicamente para el arranque inicial. Para obtener información sobre el resto de procedimientos, debe consultar el manual. Asimismo, en caso de que tuviera dudas sobre alguno de estos pasos, consulte el manual antes de continuar.
  • Seite 11 Visualización/modificación de la velocidad del agitador Puede ver/modificar el valor de la velocidad con la bañera en marcha o parada. En la pantalla de inicio, pulse la siguiente tecla multifunción para mostrar: Pulse y mantenga pulsadas las teclas de flecha para que aparezca el valor deseado. Los valores pueden ser 0 y de 30 a 200. Una vez que aparezca el valor deseado, pulse para guardar los cambios y volver a la pantalla de inicio.
  • Seite 12 Banho de água de precisão – Guia de início rápido Este guia de início rápido destina-se apenas ao arranque inicial. Para todos os outros procedimentos, consulte o manual. Além disso, se qualquer uma dessas etapas não estiver clara, consulte o manual antes de prosseguir. Preenchimento Certifique-se de que a porta de drenagem do reservatório esteja fechada e que todas as conexões do encanamento estejam firmemente sondadas ou tampadas.
  • Seite 13 Visualizar/alterar a velocidade do agitador Você pode visualizar/alterar a velocidade com o banho funcionando ou parado. Na tela inicial, pressione a tecla de função abaixo para exibir: Pressione e segure as teclas de seta para exibir o valor desejado. Os valores são 0 e 30 a 200. Quando o valor desejado for exibido, pressione para salvar a alteração e voltar à...
  • Seite 14 Guida all'avvio del bagno di precisione Questa guida di avvio rapido riguarda esclusivamente l'avviamento iniziale. Fare riferimento al manuale per tutte le altre procedure. Inoltre, qualora alcuni di questi passaggi non siano chiari, fare riferimento al manuale prima di procedere. Riempimento (Vengono filtrati solo i liquidi approvati/Acqua distillata e deionizzata) Assicurarsi che eventuali porte di scarico del serbatoio siano chiuse e che tutte le connessioni delle tubazioni siano sicure o tappate.
  • Seite 15 Visualizzazione/modifica della velocità dell'agitatore È possibile visualizzare/modificare il valore della velocità indipendentemente dal fatto che il bagno sia acceso o spento. Dalla videata iniziale, premere il tasto in basso per visualizzare: Tenere premuti tasti freccia per richiamare il valore desiderato. I valori sono 0 e 30, fino a 200.
  • Seite 16 Before calling, please obtain the following information: • bath model number • bath serial number • power source voltage The bath’s model and serial number are located on its nameplate label, see label sample on the next page. Thermo Scientific Precision...
  • Seite 17 Tel : (800) 258-0830 or Tel: (800) 258-0830 Sales: 8:00 am to 5:00 pm (603) 436-9444 Service and Support: 8:00 am to 6:00 pm Fax : (603) 436-8411 www.thermoscientific.com/tc Monday through Friday (Eastern Time) Fax: (603) 436-8411 service.tc.us@thermofisher.com Precision Thermo Scientific...
  • Seite 18 This label indicates the presence of hot surfaces. This label indicates read the manual. This label indicates a hand crushing hazard. Observe all warning labels. Never remove warning labels. Thermo Scientific Precision...
  • Seite 19 Refer service and repairs to a qualified technician. Performance of installation, operation, or maintenance procedures other than the ones described in this manual may result in a hazardous situation and will void the manufacturer's warranty and safety compliance. Precision Thermo Scientific...
  • Seite 44 General Information Chapter 2 General Information Description All Thermo Scientific Precision Water Baths provide constant temperature to applications placed in the reservoir. All have digital displays, four programmable set point temperatures, acoustic and optical alarms, and offer adjustable high temperature protection.
  • Seite 45 Specifications obtained at sea level using water. Bath depth includes bezel, height does not include lid. Thermo Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluids. Thermo Fisher Scientific reserves the right to change specifications without notice. Precision Thermo Scientific...
  • Seite 46 The installation platform must be capable of withstanding the shaking generated CAUTION when the bath is in operation. Ventilation No special ventilation clearances are required. Cover Installation Figure 3-1 – Shaking Bath Cover Installation Thermo Scientific Precision...
  • Seite 47 Once the cord is connected to the bath, connect the other end to the main power source. The following plug types are available: Bath Volts /Hertz/Phase Plug Type SWB 15 (All) 100-115/50-60/1 N5-15 200-230/50-60/1 Country Specific SWB 27 100-115/50-60/1 N5-20 200-230/50-60/1 Country Specific 1. Volts ± 10% 20 Amp Outlet 15 Amp Outlet Precision Thermo Scientific...
  • Seite 48 The usage of chlorine needs to be monitored over time in order to prevent the formation of corrosion. The duration of time that chlorine remains in solution depends on factors such as water temperature, pH and availability of direct sunlight. We recommend maintaining chlorine levels Thermo Scientific Precision...
  • Seite 49 Biocides are corrosive and can cause irreversible eye damage and skin burns. WARNING They are harmful if inhaled, swallowed or absorbed through the skin. Refer to the manufacturer's most current SDS. Precision Thermo Scientific...
  • Seite 50 Tap water is not normally recommended because the total ionized solids level may be too high. As an example, tap water in the United States averages 171 ppm (of NaCl). The recommended level for use in a water system is between 0.5 to 5.0 ppm (of NaCl). Thermo Scientific Precision...
  • Seite 51 Note Monitor the fluid level whenever heating the fluid. To conserve energy, reduce evaporation, and increase temperature control accuracy, use the supplied gable cover. Do not use aluminum foil as a cover, as it may cause corrosion due to an electrochemical reaction. Precision Thermo Scientific...
  • Seite 52 To start draining snap quick disconnect onto coupler. • To remove quick disconnect from coupler and stop draining press down on the disconnect's grey release. Coupler Quick Disconnect Neoprene Hose (3/8ID x 5/8OD) Figure 3-2 – Drain Kit Accessory Thermo Scientific Precision...
  • Seite 53 0.5 ml tubes (holds 63) Microfuge Tube Rack - 3166185 1.5 ml tubes (holds 30) Test Tube Tray 13-25mm 3161597 N.A. (holds 10) Flask Tray 25mL 3161599 N.A. (holds 18 clips) Flask Tray 50mL 3166228 N.A. (holds 10 clips) Precision Thermo Scientific...
  • Seite 54 Chapter 4 Operation Chapter 4 Operation Controller The Thermo Scientific Precision Water Baths have digital controllers that display the bath’s reservoir fluid temperature, shaker speed and other bath features. Once the circuit protector on the back of the bath is on, press to start/stop the bath.
  • Seite 55 The set point is the desired reservoir fluid temperature. 30.0°C • Press to start the bath and display the reservoir fluid temperature. If the fluid temperature is below the set point the heater will start. 29.9°C 30.0°C Precision Thermo Scientific...
  • Seite 56 T3 50.0°C 70 RPM change and return to the home screen. T4 55.0°C 90 RPM After a 30 second delay if neither key is pressed the display returns to the home screen, any change saved. Thermo Scientific Precision...
  • Seite 57 . See Chapter 6. Press, and hold, an arrow key to change the temperature. Once the desired value is displayed press to save the change or 95. °C not save the change. Press to return to the home screen. Precision Thermo Scientific...
  • Seite 58 Once the desired contrast is displayed press to return to the home screen. These icons are used to select the desired temperature scale, °C or °F. Press to toggle °C between °C and °F. Press to return to the home screen. °F Thermo Scientific Precision...
  • Seite 59 30 RPM Once the desired value is displayed press If desired, press the arrow keys to highlight another preset parameter or press to return to the home screen. Precision Thermo Scientific...
  • Seite 60 For bath shut down place the circuit protector located on the rear of the bath to the O position. • • To disconnect the bath remove the electrical plug from its power source. Thermo Scientific Precision...
  • Seite 61 Chapter 4 Operation Precision Thermo Scientific...
  • Seite 62 If corrosion (e.g., small rust marks) occur in spite of this, cleaning with stainless steel caustic agents has proved to be suitable. Apply these substances according to the manufacturer's recommendations. Electrical Power Cord Ensure any replacement cord is properly rated. Thermo Scientific Precision...
  • Seite 63 If required, use the 3/32" Allen to loosen two set screws on the eccentric and reposition the disc. Optical Disc Set Screws (located under the disc) Shaker Assembly Check the shaker assembly for looseness annually. Contact Fisher Scientific if excessive movement is noted. Precision Thermo Scientific...
  • Seite 64 HTC fault cannot be cleared, the bath must be serviced by an authorized Thermo Scientific Temperature Control Service Technician. High Temp Fault adjustable high temp fault protection limit exceeded...
  • Seite 65 Ensure bath installation complies with the site requirements in Chapter 3. • Make sure supply voltage matches bath's nameplate rating ±10%. Please contact Thermo Fisher Scientific Sales Service and Customer Support if you need any additional information, see inside cover for contact instructions. Precision Thermo Scientific...
  • Seite 67 The buyer shall be responsible for any evaluation or warranty service call (including labor charges) if no defects are found with the Thermo Scientific product. This warranty does not cover any bath that has been subject to misuse, neglect, or accident. This warranty does...
  • Seite 68 Rev 2 ECR 41929 thermoscientific.com © 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific 81 Wyman Street Waltham, MA 02451...